8 Micro Motion
®
2-Wire Coriolis Flow and Density Meters
Environmental effects
Process temperature effect Process temperature effect is defined as:
For mass flow measurement, the worst-case zero offset due to process fluid temperature
change away from the zeroing temperature.
For density measurement, the maximum measurement offset due to process fluid
temperature change away from the density calibration temperature.
Process temperature effect
% of maximum flow rate per °C density accuracy per °C
(1)
(1) For –100 °C and above.
g/cm
3
kg/m
3
CMF025 ±0,0001250 ±0,000015 ±0,015
CMF050 ±0,0001250 ±0,000015 ±0,015
CMF100 ±0,0001250 ±0,000015 ±0,015
CMF200 ±0,0005000 ±0,000015 ±0,015
Ambient temperature effect On mA output: ±0,005% of span per °C
Pressure effect Pressure effect is defined as the change in sensor flow and density sensitivity due to process
pressure change away from the calibration pressure. Pressure effect can be corrected.
Pressure effect on flow accuracy
% of rate per psi % of rate per bar
liquid gas liquid gas
CMF025 None None None None
CMF050 None None None None
CMF100 –0,0002 None –0,003 None
CMF200 –0,0008 –0,0004 –0,012 –0,006
Pressure effect on density accuracy
g/cm
3
per psi kg/m
3
per bar
CMF025 0,000004 0,058
CMF050 –0,000002 –0,029
CMF100 –0,000006 –0,087
CMF200 0,000001 0,0145
EMI effects Complies with EMC directive 2004/108/EC per EN 61326 Industrial
Conforms to NAMUR NE21 Version: 08.22.2007
Vibration limits Meets IEC 68.2.6, endurance sweep, 5 to 2000 Hz, 50 sweep cycles at 1,0 g