Instruction Manual
Appendix AX Rev. 2.1
February 1998
A-4 Appendices Rosemount Analytical Inc. A Division of Emerson Process Management
World Class 3000
TERMINAL
STRIP
REFERENCE
GAS FITTING
P00006A
CALIBRATION
GAS FITTING
VENT
Figure A-6. Junction Box
A-4 JUNCTION BOX
The junction box, Figure A-6, is positioned at
the external end of the probe and contains a
terminal strip for electrical connections and fit-
tings for reference and calibration gases. Fit-
tings are for 0.250 inch stainless steel tubing.
The calibration fitting has a seal cap which must
remain in place except during calibration. A
tubing fitting is also supplied to be used with the
calibration gas supply during calibration.
If the calibration gas bottles will be permanently
hooked up to the probe, an optional check valve
is recommended to prevent condensation of flue
gas in the calibration lines.
During operation and calibration, reference gas
is supplied through the reference gas fitting to
the reference side of the cell. This gives the
system a known quantity of oxygen with which
to compare the oxygen level in the process gas.
Unlike the non-CENELEC World Class 3000
Probe, reference air must be used in conjunc-
tion with the CENELEC World Class 3000
probe.
During calibration, two gases of different known
oxygen concentrations are injected one at a
time through the calibration gas fitting. Stainless
steel tubing delivers this gas to the process side
of the cell. In a healthy cell, the difference in
oxygen pressure from the process side to the
reference side of the cell will cause a millivolt
output proportional to the difference in oxygen
levels. The electronics unit can use the two
millivolt outputs caused by the two calibration
gases for either automatic or semi-automatic
calibration.
Do not attempt to remove a process
gas sample through either gas fitting.
Hot gases from the process would
damage gas hoses in the probe head.
A-5 CABLE ASSEMBLY
Cable used to interconnect apparatus must
conform to the applicable codes of practice in
the country of installation (example: BS4345 in
Great Britain). Rosemount can supply a cable
(P/N 1U03066) which is fitted with EExd IIC
barrier glands. The installer should note that on
some earlier versions of the cable, the glands,
although EExd IIC approved, are not of the bar-
rier gland variety. These glands should not be
used. Rosemount can supply a gland kit for the
corresponding barrier gland (P/N 1U03066G07).
Each kit contains one pair of glands. The
Rosemount supplied cable is a 7 conductor
cable to connect the probe to the HPS 3000,
and to connect the HPS 3000 to the electronics
package. Standard length for this cable is 6 m
(20 ft), but lengths up to 45 m (150 ft) are avail-
able. The 7 conductors include 1 shielded pair
of wires for the cell millivolt signal, 1 shielded
pair of type K wires for the thermocouple, and 3
individual 16-gauge wires for the heater and for
ground. All metal shields are isolated at the
probe end and connect by drain wires to ground
at the electronics. The cable is suitable for use
in ambient temperatures up to 90°C (194°F).