Viewing Battery and Monitor Status
23-21
23.14.3. Summary
String View > View|Load test or Discharge then Summary or View|Summary
Select Summary to display the Summary screen.
Figure 112. Summary (Discharges)
This screen summarizes the following information.
The String Name selected on the String Status screen.
The Starting Date and Starting Time the discharge or load test began.
The Ending Date and Ending Time the discharge or load test stopped.
The Duration of the test.
The Lowest OV (Volts) reached during the discharge or load test.
If any cells went below the threshold level during the test, Cells Below Threshold
displays a table that lists the String Number, Cell Number, and Time the cell went below
threshold.