Emerson 3350 Manual de Instruções Página 21

  • Descarregar
  • Adicionar aos meus manuais
  • Imprimir
  • Página
    / 24
  • Índice
  • MARCADORES
  • Avaliado. / 5. Com base em avaliações de clientes
Vista de página 20
20
Connect the Model 3350/3700 to a power supply as follows:
1. Use 18 to 12 AWG (0,75 to 4,0 mm
2
) wire.
2. Using a flat-head screwdriver, loosen the captive screws that secure
the display cover to the housing.
3. Ground the transmitter as follows:
Connect the ground wire to the green screw (power supply
ground; see Figure 11).
Connect the power supply ground wire directly to earth ground.
Keep all ground leads as short as possible.
Ground wiring must have less than 1 ohm impedance.
4. Connect wires to terminals 9 and 10 on the gray terminal block (see
Figures 11 and 12).
5. Close the display cover and tighten the screws.
6. A user-supplied switch may be installed in the power supply line. For
compliance with low-voltage directive 73/23/EEC (European
installations), a switch in close proximity to the Model 3350/3700 is
required.
Vista de página 20
1 2 ... 16 17 18 19 20 21 22 23 24

Comentários a estes Manuais

Sem comentários