Quick Start Guide
25
May 2015
Special Conditions for Safe Use (X):
1. If the apparatus is fitted with optional 90 V transient suppressor, it is not
capable of withstanding the 500 V insulation test required by Clause 6.3.13 of
IEC60079-11:2011. This must be taken into account when installing the
apparatus.
2. It is a condition of safe use that the above input parameters shall be taken into
account during installation.
3. It is a condition of manufacture that only the apparatus fitted with housing,
covers and sensor module housing made out of stainless steel are used in
Group I applications.
IG IECEx FISCO
Certificate: IECEx BAS 04.0017X
Standards: IEC 60079-0: 2011, IEC 60079-11: 2011
Markings: Ex ia IIC T4 Ga, T4(-60 °C
≤ T
a
≤ +70 °C)
Special Conditions for Safe Use (X):
1. The Model 3051S Transmitters fitted with transient protection are not capable
of withstanding the 500 V test as defined in Clause 6.3.13 of EN
60079-11:2012. This must be taken into account during installation.
U
i
I
i
P
i
C
i
L
i
SuperModule
30 V 300 mA 1.0 W 30 nF 0
3051S...A; 3051SF…A;
3051SAL…C
30 V 300 mA 1.0 W 12 nF 0
3051S…F; 3051SF…F
30 V 300 mA 1.3 W 0 0
3051S …A…M7, M8, or M9;
3051SF …A…M7, M8, or M9;
3051SAL…C… M7, M8, or M9
30 V 300 mA 1.0 W 12 nF 60μH
3051SAL or 3051SAM
30 V 300 mA 1.0 W 12 nF 33μH
3051SAL…M7, M8, or M9
3051SAM…M7, M8, or M9
30 V 300 mA 1.0 W 12 nF 93μH
RTD Option for 3051SF
5 V 500 mA 0.63 W N/A N/A
FISCO
Voltage U
i
17.5 V
Current I
i
380 mA
Power P
i
5.32 W
Capacitance C
i
0
Inductance L
i
0
4801_QIG_RevMA.fm Page 25 Wednesday, May 6, 2015 1:15 PM