Micro Motion
®
ELITE
®
Flow and Density Meters 25
Hazardous area classifications continued
High-temperature models CMF200A, CMF200B, CMF300A, CMF300B, CMF400A, or CMF400B with
junction box connected to MVD transmitter
Use the above graph to determine the temperature class for a given fluid and ambient temperature. The maximum surface
temperature for dust is as follows: T6:T 80 °C, T5:T 95 °C, T4:T 130 °C, T3:T 195 °C, T2:T 290 °C, T1:T 363 °C. The minimum
ambient and process fluid temperature allowed for dust is –40 °C.
The use of the sensor at an ambient temperature higher than +55°C is possible, provided that the ambient temperature does not
exceed the maximum temperature of the medium taking into account the temperature classification and the maximum operating
temperature of the sensor.
Ambient temperature range Ta –50 °C to +55 °C
Models CMF010, CMF025, CMF050, CMF100, CMF200 or CMF300 with core processor
(1)
(1) Refer to page 26 for “T” rating graph for high-temperature models with core processor.
Use the above graph to determine the temperature class for a given fluid and ambient temperature. The maximum surface
temperature for dust is as follows: T5:T 95°C, T4:T 130°C, T3:T 195°C, T2 to T1:T 254°C.
Ambient temperature range Ta –40 °C to +60 °C
Sensor fluid temp (°C)
Max ambient temp (°C)
Sensor fluid temp (°C)
Max ambient temp (°C)
Vista de página 24