Special Mounting Considerations
Mounting a Model 644H to a DIN Rail G-Rail Top Hat Rail
1. Assemble the appropriate rail clip
(part number 00644-5301-0001).
2. Attach the transmitter to a rail or panel.
3. Attach the thermowell to the pipe or process
container wall and attach sufficient lengths of
sensor lead wire. Tighten the connection
head cover.
4. Run the sensor lead wires from to the sensor
and set the transmitter failure mode switch.
5. Attach the sensor and power leads to the
transmitter.
Use with an Existing Threaded Sensor Connection Head
To mount the Models 644H and 244EH in an existing L1
connection head, assemble the Model 644H retrofit kit
(part number 00644-5321-0001) to the transmitter as
shown. Then mount the assembly in the connection
head.
644-5301B01B,D01B
Kit includes replacement
bracket and screw
644-644-5051A01A
Vista de página 62