C-6
EN
Rosemount Inc
SPECIFICATIONS
Old New
BASEFFA
(1)
Intrinsically Safe Installation Entity Parameters:
Power
Loop
U
max:in
30 V dc 30 V dc
I
max:in
200 mA 200 mA
Group II
C
Zones 0
and 1
W
max:in
1.0 W 1.0 W 0.67 W
T5 (40 T
a
40 °C)
T4 (40
T
a
80 °C)
T5 (60 T
a
40 °C)
T6 (40
T
a
40 °C)
T4 (40 T
a
80 °C)
T5 (60
T
a
50 °C)
C
eq
13.4 nF 10 nF
L
i
0 µH 0 µH
Sensor U
o
24.2 V dc 13.6 V dc
I
o
33 mA 80 mA
P
o
0.2 W 0.08 W
Cable
Load
IIC IIB IIA
IIC IIB IIA
C
o
(µF) 0.2 0.
6
1.6 0.7 5.1 18.5
L
o
(mH) 31 93 248 52348
Special Conditions for Safe Use: The apparatus
must be installed such that its external
terminals and the communication pins are
protected to at least IP20.
Special Conditions for Safe Use: The apparatus
must be installed in an enclosure which affords it a
degree of protection of at least IP20. Non-metallic
enclosures must have a surface resistance of less
than 1 G
, light alloy or zirconium enclosures must
be protected from impact and friction when
installed.