Reference Manual
00809-0600-4026, Rev BA
December 2011
Rosemount Process Radar
A-6
Echo Tuning
For more information see the sections:
“Complete the Advanced Configuration” on page 3-5 (5300),
“Verification” on page 3-7 (5300),
“Verification” on page 6-7 (5400),
“Echo-Curve Verification at Operating Conditions” on page 4-6 (5300),
“Echo-Curve Verification at Operating Conditions” on page 7-5 (5400)
and section “Configuration/Start-Up/Echo Tuning” in the Reference Manual.
Backups and Plots
For more information see the sections:
“Archive Files” on page 3-11 (5300),
“Archive Files” on page 6-10 (5400),
“Archive Files” on page 4-11 (5300) and
“Archive Files” on page 7-8 (5400).
Configuration tool used:
RRM/RCT 375/475 DD (AMS)
DTM - specify host:
Software version:
Firmware version:
Measure and learn performed?
YES NO N/A
Thresholds set at an appropriate level?
YES NO
Possible false echoes properly registered
(non-contacting only)?
YES NO
Trim Near Zone done (guided wave radar
only)?
YES NO
Probe End Projection properly activated
(guided wave radar only)?
YES NO N/A
UNZ configured?
YES NO N/A
Echo amplitude and Signal-to-Noise Ratio
OK?
YES NO
Transmitter measurement and data in the
expected range?
YES NO
Completed
Not completed
Initial backup taken and enclosed YES NO
Final backup taken and enclosed
YES NO
Initial plots taken and enclosed
YES NO
Final plots taken and enclosed
YES NO
Trends from DCS taken and enclosed
YES NO N/A
Completed
Not completed