Reference Manual
00809-0600-4026, Rev BA
December 2011
Rosemount Process Radar
5-12
6. Choose the Tank Type that corresponds to the actual tank. If none of the
available options matches the actual tank, choose Unknown.
Fast Key Sequence [1,3,4,1]
F
OUNDATION fieldbus parameter: TRANSDUCER 1100 > GEOM_TANK_TYPE
Tank Bottom Type is important for measurement performance close to
the tank bottom.
Fast Key Sequence [1,3,4,2]
F
OUNDATION fieldbus parameter: TRANSDUCER 1100 > GEOM_TANK_BOTTOM_TYPE
Tank Height is the distance from the Upper Reference Point to the tank
bottom. Make sure this number is as accurate as possible. See the
Reference Manual (Document No. 00809-0100-4026) for details.
Fast Key Sequence [1,3,4,3]
F
OUNDATION fieldbus parameter: TRANSDUCER 1100 > GEOM_TANK_HEIGHT
Comentários a estes Manuais