106 Transmitter Configuration and Use: Series 1000 and 2000 Transmitters
Pressure Compensation, Temperature Compensation, and Polling continued
9. If you will use a static pressure value:
a. Ensure that polling for pressure is not configured. Polling for temperature is allowed. See
Section 9.4.
b. Select
Static pressure.
c. Enter the static pressure.
d. Press ENTER.
e. Press SEND.
f. Press HOME.
9.3 External temperature compensation
Temperature data are used in several different calculations. Micro Motion sensors
always report temperature data to the transmitter. For greater accuracy, you can
configure the transmitter to use a different temperature value.
There are two ways to configure external temperature compensation:
If the operating temperature is a known static value, you can enter the
operating temperature in the software, and not poll a temperature
measurement device.
If the operating temperature varies significantly, you configure the
transmitter to poll for an updated temperature value from an external
temperature measurement device. Polling requires HART/Bell202
communications over the primary mA output.
Note: If your core processor is v2.1 or earlier, and you configure the transmitter for external
temperature compensation, the temperature value from the compensation procedure will replace the
sensor value in all calculations that require temperature data. If your core processor is v2.2 or later,
the temperature value from the compensation procedure is used only for enhanced density and
petroleum measurement calculations.
Note: If you configure a static temperature value, ensure that it is accurate. If you configure polling
for temperature, ensure that the external temperature measurement device is accurate and reliable.
You can configure external temperature compensation with ProLink II or the Communicator.
With ProLink II
To configure external temperature compensation with ProLink II:
1. From the
View menu, select Preferences.
2. If the
Use External Temperature checkbox does not already contain a checkmark, click the
checkbox, then click
Apply.
3. If you will poll an external device for temperature data:
a. Ensure that the transmitter is configured to use the same temperature measurement unit
that the external temperature measurement device uses (see Section 6.4).
b. Follow the polling setup instructions in Section 9.4.
Model 1500 AN
Model 1700 AN
Model 1700 IS
Model 2500 CIO
Model 2700 AN
Model 2700 IS
Model 2700 CIO