
Transmitter Configuration and Use: Series 1000 and 2000 Transmitters 169
Configuring the Enhanced Density Application continued
Enhanced Density Model 2500 CIOModel 1500 ANCalibration
D.4.2 Data points
The density curve is defined by “data points,” which are user-specified density values. You must enter
one data point for each intersection of a temperature isotherm with a concentration curve. For
example, Figure D-1 shows 5 concentration curves and 6 temperature isotherms, so 30 data points are
required to define this density curve: Point A1, Point A2, . . . Point E6.
Enter data points in the density unit that is currently configured. If the density unit is changed after
density curves have been configured, the density curves are automatically updated to use the new
density unit.
Note: See Section 6.4 for information on configuring the density unit.
The location of the data points is determined by the values used for the concentration curves and the
temperature isotherms at which the data points were measured. The range of temperature isotherms
should include the value specified for reference temperature.
D.5 Configuration procedure
Follow the instructions in this section to configure the enhanced density application.
D.5.1 Loading a standard or custom curve
Note: If the required curve is already loaded onto your transmitter, skip this step and continue with
Section D.5.2.
If a standard or custom curve has been supplied as a file, you must load it onto the transmitter.
1. Set the transmitter measurement units for density and temperature to the units used by the
curve you are loading.
• For standard curves, see Table D-3 for the units to use.
• For custom curves, see the information provided by Micro Motion.
• For information on configuring the measurement units, see Section 6.4.
2. If you are loading a standard curve, set the derived variable to Mass Conc (Dens). If you are
loading a custom curve, set the derived variable to the derived variable used by the custom
curve.
3. In the
ProLink/Configuration window, click on the ED Setup tab. The window shown in
Figure D-2 is displayed.
4. Use the
Curve being configured dropdown list to specify the slot into which the curve will be
loaded (Density Curve 1–6), and click
Apply.
5. Click the
Load this curve from a file button and specify the curve file to be loaded.
6. Repeat this process to load as many curves as required. Make sure that all loaded curves use
the same derived variable.
Note: Loading a curve that uses a different derived variable will erase all existing curve data.
Comentários a estes Manuais