Transmitter Configuration and Use: Series 1000 and 2000 Transmitters 59
Required Transmitter Configuration continued
Required Configuration Optional ConfigurationUsing the TransmitterFlowmeter Startup
Figure 6-8 Configuring the frequency output(s)
6.6.1 Configuring the process variable
With a Series 1000 transmitter, the frequency output assignment is controlled by the
primary mA output assignment. If mass flow or volume flow is assigned to the
primary mA output, the same process variable is assigned to the frequency output.
With a Series 2000 transmitter, the frequency output is independent of the primary
mA output. Table 6-10 lists the process variables that can be assigned to the
frequency output.
Note: The process variable assigned to the frequency output is always the TV (tertiary variable)
defined for HART communications. You can specify this process variable either by configuring the
frequency output or by configuring the TV (see Section 8.15.6). If you change the process variable
assigned to the mA output, the TV assignment is changed automatically, and vice versa.
If your transmitter does not have a frequency output, the TV assignment must be configured directly
(see Section 8.15.6), and the value of the TV must be queried through an RS-485 connection, read
through the Communicator, or reported through burst mode.
Table 6-10 Frequency output process variable assignments for Series 2000 transmitters
Process variable
(1)
(1) If the enhanced density application is enabled, enhanced density process variables are also available.
ProLink II code
Communicator
code Display code
Mass flow Mass Flow Mass flo MFLOW
Volume flow Vol Flow Vol flo VFLOW
Temperature-corrected (standard) volume flow
(2)
(2) Available only if the petroleum measurement application is enabled on your transmitter.
Std Vol Flow TC Vol TCVOL
Communicator DisplayProLink II
Configuration
On-Line Menu
Configure outputs
Off-line maint
Off-line configure
Configure FO
Freq/DO setup FO setup
AN and IS transmitters
CIO transmitters
Frequency/Discrete output Frequency
AN and IS transmitters CIO transmitters
ProLink Menu
Model 2500 CIO
Model 2700 AN
Model 2700 IS
Model 2700 CIO