Transmitter Configuration and Use: Series 1000 and 2000 Transmitters 119
Troubleshooting continued
Troubleshooting DiagramsDefaultsCompensation
A14 Electronics failure Transmitter Fail Cycle power to the flowmeter.
The transmitter might need service. Contact Micro
Motion. See Section 10.3.
A15 Data Loss Possible Data Loss Possible Cycle power to the flowmeter.
Perform sensor coil and resistance tests. See
Section 10.26.
The transmitter might need service. Contact Micro
Motion. See Section 10.3.
A16 Line RTD Overrange Line Temp Out-of-range Check the test points. See Section 10.24.
Check the sensor coils. See Section 10.26.
Check wiring to sensor. See Section 10.13.2.
Make sure the appropriate sensor type is
configured. See Section 6.2.
Contact Micro Motion. See Section 10.3.
A17 Meter RTD Overrange Meter Temp
Out-of-Range
Check the test points. See Section 10.24.
Check the sensor coils. See Section 10.26.
Contact Micro Motion. See Section 10.3.
A18 EEPROM Checksum—
1000/2000
EEPROM Checksum Cycle power to the flowmeter.
The transmitter might need service. Contact Micro
Motion. See Section 10.3.
A19 RAM Error—
1000/2000
RAM Error Cycle power to the flowmeter.
The transmitter might need service. Contact Micro
Motion. See Section 10.3.
A20 Calibration Factor
Unentered (Flocal)
Cal Factor Unentered Check the characterization. Specifically, verify the
FCF value. See Section 6.2.
A21 Unrecognized/
Unentered Sensor Type
(K1)
Incorrect Sensor Type Check the characterization. Specifically, verify the
K1 value. See Section 6.2.
A22 EEPROM Config
Corrupt–Core Processor
Configuration Corrupt Cycle power to the flowmeter.
The transmitter might need service. Contact Micro
Motion. See Section 10.3.
A23 EEPROM Totals
Corrupt–Core Processor
Totals Corrupt Cycle power to the flowmeter.
The transmitter might need service. Contact Micro
Motion. See Section 10.3.
A24 EEPROM Program
Corrupt–Core Processor
CP Program Corrupt Cycle power to the flowmeter.
The transmitter might need service. Contact Micro
Motion. See Section 10.3.
A25 Core Processor Boot
Sector Fault
Boot Sector Fault Cycle power to the flowmeter.
The transmitter might need service. Contact Micro
Motion. See Section 10.3.
Table 10-5 Status alarms and remedies (continued)