142 Transmitter Configuration and Use: Series 1000 and 2000 Transmitters
Default Values and Ranges continued
Density Density damping 1.6 sec 0.0 – 51.2 sec User-entered value is
corrected to nearest value in
list of preset values.
Density units g/cm
3
Density cutoff 0.2 g/cm
3
0.0 – 0.5 g/cm
3
D1 0.00000
D2 1.00000
K1 1000.00
K2 50,000.00
FD 0.00000
Temp Coefficient 4.44
Slug flow Slug flow low limit 0.0 g/cm
3
0.0 – 10.0 g/cm
3
Slug flow high limit 5.0 g/cm
3
0.0 – 10.0 g/cm
3
Slug duration 0.0 sec 0.0 – 60.0 sec
Temperature Temperature damping 4.8 sec 0.0 – 38.4 sec User-entered value is
corrected to nearest lower
value in list of preset values.
Temperature units Deg C
Temperature calibration factor 1.00000T0.0000
Pressure Pressure units PSI
Flow factor 0.00000
Density factor 0.00000
Cal pressure 0.00000
T-Series sensor D3 0.00000
D4 0.00000
K3 0.00000
K4 0.00000
FTG 0.00000
FFQ 0.00000
DTG 0.00000
DFQ1 0.00000
DFQ2 0.00000
Special units Base mass unit g
Base mass time sec
Mass flow conversion factor 1.00000
Base volume unit L
Base volume time sec
Volume flow conversion factor 1.00000
Event 1 Variable Density
Type Low alarm
Setpoint 0.0
Setpoint units g/cm
3
Table A-1 Transmitter default values and ranges (continued)
Type Setting Default Range Comments